Informations sur le produit "Fluorescence/UV Quartz Cuvette"
Designed specifically for fluorescence measurement, this 10 x 10 mm cuvette is made of quartz and features four clear sides to enable crystal-clear transmission of light, even into the UV spectrum.
Bildungsstufe:
Sec. II & tertiaire, Secondaire II
Fach:
biologie, chimie
Fachbereich:
Biologie, Chemie
0 sur 0 évaluations
Note moyenne de 0 sur 5 étoiles
Laissez une évaluation !
Partagez avec d'autres clients votre avis sur le produit.
Connexion
Could not translate review. Please try again later
Aucun avis n'a été trouvé. Partagez vos idées avec d'autres personnes.
The affordable Fluorescence/UV-VIS Spectrophotometer is designed specifically for upper-level chemistry courses. Vernier International has combined the functionality of a fluorometer with the power of a UV/VIS spectrophotometer to allow students to easily and accurately conduct numerous fluorescence experiments with compounds such as quinine sulfate, DAPI, GFP, and tryptophan. Exchangeable LEDs ensure you get the exact excitation wavelength needed for your experiment. To conduct emission spectra experiments simply attach the Vernier Spectrophotometer Optical Fiber.
The Fluorescence/UV-VIS Spectrophotometer connects directly to a computer or LabQuest. Data can be collected and analyzed using our Logger Pro software, so there is no need to train students on yet another instrument software.
Specifications
Detection modes: Absorbance and Fluorescence
Dimensions: 18.5 cm × 17.0 cm × 7.0 cm
Power supply: AC adapter (included)
Power consumption: 3 A start-up, 500 mA continuous
Absorbance light source: deuterium (UV) and incandescent (VIS)
Emission light source: Exchangeable LED (ships with 375 nm, 450 nm, 525 nm)
Detector: Linear CCD
Wavelength range: 220–850 nm
Wavelength reporting interval: ~ 1 nm
Optical resolution: 3.0 nm (as determined with 486 nm hydrogen emission spectral line FWHM)
Wavelength accuracy: ± 2.0 nm (as determined with holmium oxide NIST standard)
Absorbance photometric accuracy: ± 5.0% (as determined with potassium dichromate NIST standards)
Absorbance photometric range (for best accuracy): 0.1–1.0
Typical scan time: ~ 2 s
Sample format: 10 mm × 10 mm cuvette (UV fluorescence cuvette included)
Fluorescence emission detection limit: 1 mg/L quinine sulfate dihydrate in 0.1 M H2SO4
Included
Vernier Fluorescence/UV-VIS Spectrophotometer with LED holder
Three exchangeable LED cartridges: 375 nm, 450 nm, and 525 nm
Starna UV fluorescence cuvette
AC power supply
USB cable
Carrying case
For use with the Fluorescence/UV-VIS Spectrophotometer, this filter acts as an excitation source to cause the sample to fluoresce. Because the excitation source is an exchangeable LED, not only do you get the standard functionality of a spectrofluorometer but software controls allow students to determine the LED intensity and explore fluorescence as a function of excitation.
Available LED Cartridges:
VSP 280 nm LED
VSP 350 nm LED
VSP 375 nm LED
VSP 400 nm LED
VSP 450 nm LED
VSP 500 nm LED
VSP 525 nm LED
Un paquet de 100x 3,5 ml (cuvettes avec 20 couvercles) est disponible pour une utilisation avec notre colorimètre (COL-BTA), SpectroVis Plus (SVIS-PL), et Vernier Spectrometer (V-SPEC).