The O2 Gas Sensor can be used to monitor gaseous oxygen levels in a variety of biology and chemistry experiments. The sensor is intended for measuring gaseous, not aqueous, O2 concentration.

Specifications

  • Measurement Range: 0–27% (0-270 ppt)
  • Accuracy (@ Standard Pressure 760 mm Hg): +/- 1% O2
  • Typical Resolution: 0.01%
  • Response Time: 90% of final reading in ~12 seconds
  • Warm-Up Time: Less than 5 seconds
  • Pressure Effect:
  • Directly proportional
  • Vout = Vout (standard) x (P/1013)
  • Pressure range: 0.5 atm to 1.5 atm
  • Output Signal Range: 0.3 to 5 VDC; 2.7 to 3.8 VDC @ 21% O2
  • Output Impedance: 1 KΩ
  • Input Voltage: 5 VDC +/–0.25 VDC
  • Gas Sampling Mode: Diffusion
  • Normal Operating Temperature Range: 20°C to 30°C
  • Operating Temperature Range: 5 to 40°C (with calibration)
  • Operating Humidity Range: 0 to 95% RH
  • Storage Temperature Range: –20 to +60°C
  • Dimensions:
  • Sensor Tube: 45 mm length, 28 mm OD
  • Total length: 135 mm

What's included

  • Vernier O2 Gas Sensor
  • 250 mL gas sampling bottle

User Manual - O2-BTA

Bildungsstufe: Diploma/ Vocational/ Uni, Middle years
Fach: Biology, Vernier, chemistry

0 of 0 reviews

Average rating of 0 out of 5 stars

Leave a review!

Share your experiences with other customers.


Skip product gallery

Accessory Items

BioChamber 2000
The BioChamber 2000 is a 2000 mL plastic chamber that can be used with a CO2 Gas Sensor and an O2 Gas Sensor to monitor gaseous carbon dioxide and oxygen levels simultaneously in a closed system. User Manual - BC-2000

CHF 50.90*
Human Physiology with Vernier
The 24 experiments in Human Physiology with Vernier are designed to encourage students to think about the physiology of various human organ systems by measuring the activity of the upper respiratory tract, the lungs, the heart, the nervous system, and the musculoskeletal system. Homeostasis is an underlying theme, and several experiments illustrate the response of the human body to stress or external influences.

From CHF 83.40*
O2 REPLACE SENSOR+LABEL-FOR INTL DEALERS ONLY

CHF 130.00*
BioChamber 250
The BioChamber 250 is a 250 mL Nalgene bottle that can be used with a CO2 Gas Sensor and an O2 Gas Sensor to monitor gaseous carbon dioxide and oxygen levels simultaneously in a closed system. User Manual - BC-250

CHF 18.30*
250 mL Nalgene Bottle w/Lid
This is the same bottle that ships with the CO2 Gas, O2 Gas, and Ethanol Sensors. User Manual - CO2-BTL

CHF 10.10*
O2 Gas Sensor to Spirometer Adapter
This adapter facilitates the connection between an oxygen gas sensor and a spirometer when measuring the oxygen gas concentration and flow rate of exhaled air. It is used in experiment #22: Oxygen and Aerobic Metabolism from Human Physiology with Vernier. User Manual - O2-SPR

CHF 18.30*
Skip product gallery

Similar Items

O2 Gas Sensor
The Go Direct O2 Gas Sensor measures gaseous oxygen concentration levels and air temperature. It is useful for a variety of biology and physiology experiments. It connects via Bluetooth® wireless technology or via USB to your device. Specifications Oxygen SensorCell type: Electrochemical cell Range: 0–100% (0–1000 ppt) O2 Accuracy (at standard pressure 760 mm Hg): ±1% O2 Resolution: 0.01% O2 Temperature SensorType: Thermistor Accuracy: ±0.5°C Resolution: 0.1°C DimensionsSensor tube length: 38 mm Sensor tube diameter: 28 mm Total length: 155 mm ConnectionsWireless: Bluetooth Wired: USB What's included Go Direct® O2 Gas Sensor Micro USB Cable 250 mL Nalgene Bottle w/Lid User Manual - GDX-O2

CHF 487.20*